Designed specifically for Hitachi specimen holders, the ZONESEMⅡ provides optimized sample cleaning and storage modes of operation. The ZONESEMⅡ serves as an excellent desiccator to prevent holder out-gassing.
|
Interactions between hydrocarbons on the specimen surface and the irradiated electron beam generate sample contamination. Hydrocarbons are generally non-covalently attached to the specimen surface during sample handling, preparation, and storage.
The ZONESEMⅡ uses a UV lamp with wavelengths of 185 nm and 254 nm to efficiently remove hydrocarbons. Ozone is generated from oxygen in the vacuum-controlled specimen chamber, and generated ozone is broken into oxygen and activated oxygen. Hydrocarbons on the specimen surface are broken into volatile molecules such as CO2, H2O by radical oxygen and UV. The exhaust pump evacuates the free molecules from the sample chamber.
Without “ZONESEMⅡ” cleaning |
After “ZONESEMⅡ” cleaning |
Without “ZONESEMⅡ” cleaning | After “ZONESEMⅡ” cleaning |
ZONESEMⅡ enhances EDS analysis.
SEM Image
EDS Mapping Image
Without “ZONESEMⅡ” cleaning |
After “ZONESEMⅡ” cleaning |
Description | |
---|---|
Model No. | ZONE-2010 |
UV Lamp | Wavelengths: 184 nm, 254 nm
(visible light, green and infrared light) |
Specimen Size | 100 mm(D) × 37 mm(H) |
Process Mode | UV Cleaning mode, Vacuum storage mode |
Vacuum Setting | Settable by 100 phases of vacuum pressure |
Process Setting Time | One minute to approximately 24 hours |
Vacuum Pump | Dry vacuum exhaust system |
Ozone exclusion | Built-in Ozone exclusion device |
Unit Dimension | 310 mm(W) × 260 mm(D) × 305 mm(H) |
Weight | 18 kg |
Setting Condition
Description | |
---|---|
Room Temperature | 15 - 30 degrees C |
Humidity | ≤ 70% (no condensing dew) |
Power Supply | Power Supply: AC100 V - 240 V(±10%), 50/60 Hz 0.14 kVA |