System electrical requirements:
100/120 Volts (50-60 Hz)
220/240 Volts (50-60 Hz)
Power consumed:
At 115 V.A.C.: Controller: 72 Watts typical; Incubator: 640 Watts typical; Low Temperature Cart: 660 Watts typical
At 230 V.A.C.: Controller: 72 Watts typical; Incubator: 640 Watts typical; Low Temperature Cart: 605 Watts typical
Heat dissipated:
Controller: 245 BTU/hr, 2840 maximum
Incubator: 904 BTU/hr, 2840 maximum
Low Temperature Cart: 2249 BTU/hr at 660 watts
Low Temperature Cart: 2061 BTU/hr at 605 watts
Controller Module dimensions:
Width: 35.6 cm (14 in)
Height: 91.4 cm (36 in)
Depth: 61.7 cm (24.3 in)
Weight: 57.2 kg (126 lbs)
Incubator Module dimensions:
Width: 49.6 cm (19.5 in)
Height: 91.4 cm (36 in)
Depth: 61.7 cm (24.3 in)
Weight unloaded: 118.8 kg (262 lbs)
Weight loaded: 133.8 kg (295 lbs)
Low Temperature Module dimensions:
Width: 80.9 cm (31.9 in)
Height: 184.6 cm (72.7 in)
Depth: 66.1 cm (26.0 in)
Weight unloaded: 293.5 kg (647 lbs)
Weight loaded: 313.9 kg (692 lbs)
Benchtop cart dimensions:
Width: 99.1 cm (39 in)
Height: 85.1 cm (33.5 in)
Depth: 63.5 cm (25 in)
Weight: 100 kg (220 lbs)
System operating and storage humidity range:
Storage: 20-85% relative humidity, non-condensing
Operating: 20-60% relative humidity, non-condensing
System operating and storage altitude:
6562 feet (2000 meters)