● High-sensitivity, high-precision measurement by the differential method
● Multi-measurement system that excels in expandability
● Flexible handling of various sample sizes
● Notable enhancement of heating and cooling rates enabled by a compact electric furnace
● Simple sample setting mechanism, the first of its kind in differential TMA
● Safety foremost emphasized on the entire system
● Different measurement methods
+ Compression loading method
+ Tensile loading method
+ Penetration method
+ High-sensitivity differential penetration method
● Measuring temp. range
+ Standard model: Ambient to 1100°
+ High-temperature model: Ambient to 1500°
Measurement mode | Compression loading*1 | Tensile loading | Penetration | ||
Type | Standard | High temperature | — | Penetration | High sensitivity differential penetration method |
Detection system | Differential expansion | Non-differential expansion | Differential expansion | ||
Standard sample size | 5 mm diameter (max 9 mm diameter) | Thickness: 10 - 200 µm | 5 mm diameter | ||
Length: 10 - 20 mm | Length: 10, 15, 20 mm; width: 5 mm | Thickness: 4 mm (max) | |||
Material | SiO2 | Al2O3 | SiO2 | ||
Maximum load | 1000 mN | ||||
Measurement temperature range*2 | Ambient to 1100°C | Ambient to 1500°C | Ambient to 600°C | ||
Maximum heating rate | 100°C/min (low temperature furnace: 20°C/min) | ||||
Measurement range (FS) | 5000 µm | ||||
Loading mode |
|
||||
Measurement atmosphere | Air, inert gas or gas flow |
*1 The specifications of the high sensitivity three-point bending method conform with the standard model for compression loading mode
*2 Option: low temperature furnace: -150°C to 600°C; Liquid nitrogen auto feed type cooling attachment: -150°C to 600°C
1. Measurement data protection function
During measurement, when a connection error occurs between the station and the module, the module will continue to perform
the on-going measurement and the data will be stored in the module. The stored data is retrieved using a PC after the measurement,
protecting valuable measurement data.
2. End operation settings function
The ECO mode or power OFF can be selected after the end of measurement. The temperature and time condition can be specified.
We can select a variety of operation condition of the attached forced air-cooling fan such as cool up to room temperature, not in use
on continuous usage.
3. ECO mode
After completing measurement, the ECO mode can be selected in the stand-by condition reducing the electricity consumption.
Especially in TG-DTA, the ECO mode facilitates immediate transition from stand-by to stable measurement conditions.
4. First aid function
When an error or problem occurs in the module, the error number, error description as well as troubleshooting measures will be displayed
facilitating a quick and smooth recovery condition. Also the error contents are stored in log files and can be readily sent to the technical service
department for an immediate appropriate response.
5. ASCII data import function
Using the conversion software, the Thermo plus EVO2 analysis software can analyze carious ASCII converted measurement data.
6. E-mail function
The Thermo plus EVO2 can transmit information by email such as end of measurement, measurement data, occurrence of an error to PC,
mobile phones thru corporate LAN connection abs can confirm the condition of an on-going measurement at remote locations.
7. Excel, Word output
From the menu directly, the measurement file can be directly exported in Word or Excel. Also, the output style can be modified thru the layout
editing function. When exporting data in Excel, the numerical data will be automatically created in the worksheet 2 allowing a convenient data
analysis in other application software.
8. Instrument usage history listing function
This function automatically records the module’s usage. Information such as date of usage, module name, operator, temperature program and
measurement results are filed and saved in the achieve list. The history use of the module and the time of use are clear and effective for
management and maintenance.
Hiện tại chưa có ý kiến đánh giá nào về sản phẩm. Hãy là người đầu tiên chia sẻ cảm nhận của bạn.