Model
|
VM-1970
|
Input Signal
|
Piezoelectric Accelerometer
|
Frequency Range
|
Acceleration: 1Hz ~ 100kHz (±3dB), 3Hz ~ 70kHz (±0.5dB)
Velocity: 3Hz ~ 6kHz (±0.5dB)
Displacement: 3Hz ~ 600Hz (±0.5dB)
|
Measurement Range (for an Accelerometer of Charge Sensitivity 1.0~9.99pc / m/s2)
|
Acceleration
|
1,10,100,1000,10000m/s2,5-Ranges
|
Velocity
|
3Hz~ 1,10,100,1000 cm/s 4-Ranges
10Hz~ 1,10,100,1000 cm/s 4-Ranges
30Hz~ 0.1,1,10,100,1000 cm/s 5-Ranges
|
Displacement
|
3Hz~ 1,10,100,1000 mm p-p 4-Ranges
10Hz~ 0.1,1,10,100,1000 mm p-p 5-Ranges
30Hz~ 0.01,0.1, 1,10,100 mm p-p 5-Ranges
|
H Function
|
Value of Acceleration passed through 1kHz Low-pass Filter after processed by the 2kHz~15kHz Band-pass Filter. Unit is same as of Acceleration.
|
|
Filter Response
|
Lower Cutoff Frequencyr: (LOW.F):OFF (1Hz),3,10,30Hz
Higher Cutoff Frequency: (HIGH.F):300Hz,1,3,10kHz, OFF,(100kHz)
Cutoff characteristic: -18dB, Butterworth
|
Indication
|
rms : Mean Square value EQP: Equivalent Peak estimated from average value PEAK: Peak value P・HOLD: Held Max. Peak Value
|
Level Watching Function
|
Alarm Level Step: 1 Step foe every Channel Setting: 0 ~110% for every range Alarm Output: 1a Relay Contact Output Alarm Output : 1a Relay Contact shared by every channel ITTL Level Output for each individual channel Contact Capacity: DC3VDC 0.5A, AC120V 0.5A
|
Output
|
Waveform: Voltage: (Load 10kΩ or higher)
Level Output: Voltage: (Load 10kΩ or higher)
Calibration Output: 80Hz Sine
|
Ambient Conditions
|
For Operation: 0 ~ 40°C, 0 ~ 85% RH or lower
|
Power
|
100V AC ±15%, 50/60Hz, ≦20VA
Battery: D size 4 pcs. Life 10 hours or longer
External: DC9~12V
|
Main Unit Size・Mass
|
200Wx150Hx290Dmm (excluding Handles and other protrusions) Approx. 5kg
|