● Microprocessor PID control.
● Forced air convection using an optimal cross flow fan for chamber temperature uniformity.
Model |
LCH-11 |
LCH-21 |
LCH-31 |
LCH-11- |
LCH-11G |
LCH-21G |
LCH-31G |
LCH-11G- |
|
Chamber Volume (L / cu ft) |
150/5.3 |
255/9.0 |
485/17.1 |
7150 |
150/5.3 |
255/9.0 |
485/17.1 |
150 |
|
Temperature |
Range |
-20 to 100 / -4 to 212 |
0 to 100 / 32 to 212 |
||||||
Fluctuation (±℃ / ℉) |
0.2 to 0.3 / 0.36 to 0.54 |
||||||||
Variation at 80℃ (±℃ / ℉) |
0.8 / 1.44 |
||||||||
Time |
Heating (min.) |
80 (from 20℃ to100℃) |
|||||||
Cooling (min.) |
60 (from100℃ to 40℃) |
||||||||
Dimension |
Interior (mm / inch) |
600x500x500 |
600x500x850 |
700x680x1020 |
600x500x500 |
600x500x500 |
600x500x850 |
700x680x1020 |
600x500x500 |
Exterior (mm / inch) |
794x1070x1280 |
794x1070x1630 |
894x1250x1800 |
794x1070x1900 |
794x1070x1280 |
794x1070x1630 |
894x1250x1800 |
794x1070x1900 |
|
Net Weight (kg / lbs) |
125/275.6 |
185/407.9 |
225/496.0 |
195/429.9 |
120/264.6 |
180/396.8 |
220/485.0 |
190/418.9 |
|
Cable Port (mm / inch) |
Ø 80 / Ø 3.15 |
||||||||
Interface Port |
RS-232C |
||||||||
Electrical Requirements (230V, 60Hz) |
6.3A |
9.2A |
11.1A |
12.6A |
7.1A |
10.6A |
13.8A |
14.2A |
|
Cat. No. |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
|
Electrical Requirements (230V, 50Hz) |
6.6A |
9.6A |
11.6A |
13.2A |
7.4A |
11.1A |
14.4A |
14.8A |
|
Cat. No. |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
AAHK |
|
Electrical Requirements (120V,60Hz) |
11.6A |
23.2A |
13A |
26A |
|||||
Cat. No. |
AAHK |
- |
- |
AAHK |
AAHK |
- |
- |
AAHK |
Hiện tại chưa có ý kiến đánh giá nào về sản phẩm. Hãy là người đầu tiên chia sẻ cảm nhận của bạn.