1. Introduction
120 kV transmission electron microscopes (TEMs) are used in a wide variety of fields, from biomedical research to nanomaterial science, to observe and analyze the fine structure and morphology of samples. At Hitachi we have developed the HT7800 series of 120 kV TEM instruments to address this broad range of applications 1 .
The HT7800, designed for biomedical research, is equipped with a high-contrast lens to enable high-contrast observations over a wide field of view. The HT7800 also features acquisition of 3D image data with sample tilt angle variable over a range of ±70 degrees and 3D reconstruction function as a standard.
The HT7830, designed for nanomaterial science, achieves TEM resolution of 0.19 nm (on-axis) and STEM resolution of 1 nm. It includes Nano Analysis2)function for automatic a cquisition of multiple d iffraction patterns in selected areas.
For field-of-view searching, both the HT7800 and HT7830 both include Image Navigation3, a feature that offers intuitive field-of-view searching with assignment of areas in images and automatic image acquisition.
Now Hitachi has developed the HT7820, an instrument with specifications intermediate between those of the HT7800 and HT7830 and featuring a multi-purpose objective lens to meet the observational needs of both biomedical researchers and nanomaterial scientists.
In this article, we discuss the features of the HT7820—including common functionality shared by all instruments in the HT7800 series—and some applications.
2. Overview and Key Features of the HT7820
Figure 1 shows the external view of HT7820 120 kV TEM. The HT7800 series including the HT7820 uses a screen camera for fluorescent-plate observations to enable TEM operation in bright rooms. The high-definition, high-speed CMOS screen camera provides rapid field-of-view searching with minimal beam damage to a sample4). Moreover it enables live measurements of the image, recording images and movies on the fluorescent plate.
As the predecessors in the HT7800 series, the HT7820 boasts a compact design concept with a space-saving footprint [with main unit dimensions of 199 cm (H) × 191 cm (D) × 120 cm (W)] and reduced power consumption.
Table 1 shows main specifications of HT7800 series. All instruments in the series, including the HT7820, allow the accelerating voltage ranging from a minimum of 20 kV to a maximum of 120 kV. This allows the accelerating voltage to be chosen appropriately for the purpose of observation, from high-contrast observations of unstained sample fragments to low-damage observations of carbon-based materials.
HT7800 | HT7820 | HT7830 | |
Accelerating voltage | 20 - 120 kV (100 V/step variable) | ||
Lattice resolution | 0.204 nm (Off-axis, 100 kV) |
0.14 nm (Off-axis, 120 kV) |
0.14 nm (Off-axis, 120 kV) 0.19 nm (On-axis, 120 kV) |
Magnification | x600,000 | x800,000 | x1,000,000 |
Maximum tilt angle (standard sample holder) |
±70° | ±30° | ±10° |
Lens type |
Dual-mode objective lens |
||
High-contrast observations over |
Multi-purpose observations | High-resolution observations |
Table 1: Main specifications of HT7800 series instruments
The HT7820 guarantees lattice resolution of 0.14 nm (off-axis) and a sample tilt range of ±30 degrees, primarily for research in the fields of semiconductors and soft matters such as polymers and light-weight materials. Figure 2 shows an example of a high-resolution image observation of single-crystal gold (Au); the 0.14 nm lattice spacing of the Au (220) is clearly observed.
Fig. 2: High-resolution image observation of a single-crystal gold sample
The HT7820 is equipped with Hitachi’s unique dual-mode objective lens, which allows to switch between highcontrast (HC) mode and high-resolution (HR) mode depending on the purpose of observations5). As illustrated in the schematic diagram of Figure 3, the dual-mode objective lens consists of two-stages in which the objective magnetic path exists in a double-gap configuration; the OBJ2 (lower) lens may be turned on or off by switching the polarity of its coil, allowing both long (HC mode) and short (HR mode) focal length to be realized by a single lens assembly. The advantage of this design is that the transition between HC and HR modes requires switching only the polarity of the coil, with the magnitude of the current remaining constant to prevent thermal drift due to temperature variation.
Fig. 3: Schematic diagram of the dual-mode objective lens
In addition, the HT7820 includes a variety of automated functional tools—such as Image Navigation and Nano Analysis—and is compatible with Hitachi’s RC16 16-mega pixel CMOS camera and many other types of cameras.
3. Applications of the HT7820
Figures 4 and 5 show examples of high-contrast image observations. The sample in Figure 4 is an impact-resistant polystyrene material, whose salami-like structure is captured by the high-contrast image observation. The sample in Figure 5 is an unstained section of a mouse kidney; although the sample is unstained, the use of HC mode allows highcontrast observation of cellular structures such as the glomerulus nucleus and the podocyte foot processes.
Fig. 4: High-contrast image observation of impact-resistant polystyrene sample |
Fig 5: High-contrast image observation of unstained mouse kidney sample |
3-2. High-resolution image observations
Figure 6 shows a high-resolution image of a silicon (Si) single-crystal [Figure 6 (a)] and the corresponding diffraction pattern [Figure 6 (b)]. Lattice images are clearly discernible in three directions; this capability is useful for calibration of length measurements in Si semiconductor devices. Figure 7 shows a high resolution image of a TiO2 nanoparticle, in which 0.35 nm lattice images are clearly observed.
Fig 6: High-resolution image (a) of a Si single-crystal and the corresponding diffraction pattern (b) |
Fig 7: High-resolution image of TiO2 nanoparticle |
3-3. TEM images obtained at different accelerating voltages
Figure 8 shows images of a single-walled carbon nanotube observed at accelerating voltages ranging from 20 kV to 120 kV in steps of 20 kV. The use of lower accelerating voltage allows individual nanotubes to be discerned with higher contrast.
Fig 8. Single-walled carbon nanotube sample observed at different accelerating voltages
3-4. Automatic image acquisition using Image Navigation
Figure 9 illustrates the Image Navigation function for automated image acquisition. The sample is asbestos. First, a grid view is displayed on the navigation screen by a single clicking [Figure 9 (a)]. After the target area is selected, images of the target area (b) at the specified magnification (c) are captured automatically (d). This allows automated image capture for all grid sectors in which asbestos is present. The automatically-captured images are recorded to the data base with image information and displayed in a thumbnail view. The Auto drive function for moving the stage may be invoked by selecting an image corresponding to the desired field of view (e). Subsequently, the imaging process may be further automated by selecting multiple desired fields of view and specifying an imaging magnification.
Fig 9: Automated image acquisition for an asbestos sample using Image Navigation function
3-5. Nano Automatic acquisition of selected-area diffraction (SAD) patterns using Nano Analysis function
Figure 10 shows an example of automated acquisition of selected-area diffraction (SAD) patterns using the Nano Analysis function, in this case for a sample of asbestos. In TEM image, 10 regions, indicated by circles corresponding to the aperture diameter, are selected. Then the appropriate selected area aperture is inserted into the center of the screen.
Fig 10: Automatic acquisition of selected-area diffraction (SAD) patterns in an asbestos sample using Nano Analysis function
By starting automatic SA-diff acquisition, the field of view and the corresponding diffraction pattern are recorded for each selected region.
Furthermore, combining with the Diffraction analysis function, matching candidates of material can be listed up from the database.
Figure 11 shows TEM images, SAD patterns, and EDX analysis results acquired by the HT7820 for asbestos samples. The addition of EDX analysis ensures the identification of asbestos.
Fig 11: TEM images, SAD patterns, and EDX analysis results acquired by the HT7820 for asbestos samples
3-6. Multi-directional observations with the grid-rotation holder option6
The grid-rotation holder allows 360-degree rotation of samples mounted in a rectangular grid for TEM observations. The use of this holder allows acquisition of three-dimensional tilt images and reconstruction—even on instruments with narrow-gap objective lenses, such as the HT7820 and HT7830.
Figure 12 shows the use of this technique to obtain 3-directional images of a sample of molybdenum oxide on a holy carbon support.
Fig 12. Multidirectional observation using the grid-rotation holder
The left, center, and right panels show TEM images and SAD patterns captured at sample tilt angles of -55, 0, and +55 degrees, respectively. The multi-directional observation yields information on the morphology of the molybdenum oxide crystal and crystal orientation from SAD patterns.
With a single observed image (i.e, for tilt angle -55 degrees), it would be concluded erroneously that two types of crystal morphology—one needle-shaped and one planar—were present. The multi-directional observation clearly identifies that both crystals are in fact planar. This demonstrates the possibility of accurately determining sample morphology from just a few observations at various tilt angles. The accuracy can be further improved by increasing the number of tilt angles to produce a three-dimensional reconstruction, yielding high-precision information on crystal size and morphology.
4. Conclusions
In this article we discussed the features of Hitachi’s HT7800 series—with a particular focus on the HT7820, the most recent addition to the lineup—and presented case studies of these instruments to analyze a variety of materials. The HT7820 is a versatile tool whose wide range of applications facilitates the screening and testing of many types of samples.
References
1)I. Nagaoki et al., THE HITACHI SCIENTIFIC INSTRUMENT NEWS, 10, 18–21 (2018).
2)K. Tamura et al., Microsc. Microanal, 23 (Suppl 1), 62–63 (2017). 3)H. Kawamoto et al., Proc. of IMC 19, Sydney, Australia, 228 (2018).
4)T. Yaguchi et al., Abstracts of the 74th Annual Meeting of the Japanese Society of Microscopy, 35 (2018) (in Japanese). 5)K. Tamura et al., Abstracts of the 74th Annual Meeting of the Japanese Society of Microscopy, 4 (2018) (in Japanese). 6)A. Wakui et al., Abstracts of the 74th Annual Meeting of the Japanese Society of Microscopy, 217 (2018) (in Japanese).
About the authors
*1 Toshie Yaguchi
Analysis Systems Design 2nd Department Metrology and Analysis Product Division Nano-Technology Solution Business Group Hitachi High-Technologies Corporation
*2 Mami Komomi
Analysis Systems Solution Development Department Metrology and Analysis Product Division
Nano-Technology Solution Business Group Hitachi High-Technologies Corporation