Application Note - Digestion of Beef with Nano particles

1641 views - 29-07-2019, 9:07 am

Introduction:

This study evaluated the effectiveness of the NovaWAVE for the digestion of beef spiked with TiO2.

 

Supplies and Reagents:

1) NovaWAVE SA Model*

2) Teflon® 75 ml with Quartz Sleeve Vessels*
3) Teflon® Caps and Safety Pressure Release Caps (pre-set release pressure to 435 psi)*
4) PlasmaPURE HNO3 (70%)
5) PlasmaPURE HF (48%)
6) ICP-OES, Spectroflame FMD-007, Spectro ®

* Manufactured by SCP SCIENCE

 

Sample Preparation Procedure:

The beef was weighed (0.2) on a 4 place analytical balance directly into the 75 ml Teflon vessels. After adding 4:1 HNO3:HF (5 ml total) of acid and 0.01 g of TiO2, the samples were digested in the NovaWAVE following the instrument’s heating profile noted below. After cooling, the samples were diluted to 200 ml with water in a volumetric flask. The solutions are clear with no particulates remaining.

 

NovaWAVE Heating Program:

Stage

Ramp time

(minutes) 

Pressure

(PSI – Limit)

Temperature

(oC) 

Hold time

(minutes)

1 10 435 200 15

 

Analysis:

Element

Beef

(ppm)

Spike TiO2

(ppm)

Beef + TiO2

(ppm)

Spike recovery

(%)

Ti 8.53 12000 12002 100

Conclusion

 

The sample had to made from 2 components as it is very difficult to mix TiOinto ground beef to ensure homogeneity. NovaWAVE work well to digest a sample of meat that has Nano particles in the sample.

 

******************************************