A quick introduction to the AP-5000
The AP-5000 features 4 Auxiliary sockets to allow you to add further sensors to the probe. Unlike the AP-2000, these auxiliary ports are unrestricted. This means that each port can take either an Ion Selective Electrode (ISE) or an optical sensor. This makes the number of potential sensor combinations vast.
The probe can be used with a variety of cable lengths via the AquaConn Connectors. Cables available in 3, 10, 20 and 30m lengths as standard.
Customising the AP-5000 is simple, there are 4 ports and you can add any combination of our ISE and optical sensors into these ports.
Probes are supplied with blanking plugs in the auxiliary ports as standard. Simply unscrew these to add your extra electrodes.
The AP-5000 features a depth sensor. This sensor is capable of measurement accuracy of 0.5%FS. This is very comparable to a dedicated water level logger such as the LeveLine-ECO: 0.2%FS, and the LeveLine (titanium): 0.1%FS.
The depth is zeroed automatically whilst the conductivity is reading zero. As soon as the conductivity registers over 0 the probe depth begins to be calculated. Once the probe is lowered to the desired depth it must be secured, then using the GPS Aquameter the depth can be zeroed once again. Now the AP-5000 will detect small changes in the water level and display them as a positive or negative change.
The optical sensor works on the principle of Dynamic Luminescence Quenching. A gas-permeable material known as a luminophore is excited with short bursts of blue light, which causes molecules in the luminophore to emit red photons. By measuring the delay of the returned red photons with respect to the blue excitation, the level of dissolved oxygen present can be determined. The optical method has various advantages over the historical galvanic method for measuring dissolved oxygen.
The most important being that as no oxygen is consumed across a membrane, the sensor does not require a flow of liquid passing over it to achieve a stable reading. Other advantages include in frequent calibrations (every 3-6 months), and replaceable caps that last over 2 years.
Four Auxiliary Sockets
Dissolved Oxygen | Range | 0 – 500.0% / 0 – 50.00 mg/L |
Resolution | 0.1% / 0.01mg/L | |
Accuracy | 0 – 200%: ± 1% of reading. 200% – 500%: ± 10% | |
Conductivity (EC) | Range | 0 – 200 mS/cm (0 – 200,000 µS/cm) |
Resolution | 3 Auto-range scales: 0 – 9999 µS/cm, 10.00 – 99.99 mS/cm, 100.0 – 200.0mS/cm | |
Accuracy | ± 1% of reading | |
TDS* | Range | 0 – 100,000 mg/L (ppm) |
Resolution | 2 Auto-range scales: 0 – 9999mg/L, 10.00 – 100.00g/L | |
Accuracy | ± 1% of reading | |
Resistivity* | Range | 5 Ωcm – 1 MΩcm |
Resolution | 2 Auto-range scales: 5 – 9999 Ωcm, 10.0 – 1000.0 KΩcm | |
Accuracy | ± 1% of reading | |
Salinity* | Range | 0 – 70 PSU / 0 – 70.00 ppt (g/Kg) |
Resolution | 0.01 PSU / 0.01 ppt | |
Accuracy | ± 1% of reading | |
Seawater Specific Gravity* | Range | 0 – 50 st |
Resolution | 0.1 st | |
Accuracy | ± 1.0 st | |
pH | Range | 0 – 14 pH / ± 625mV |
Resolution | 0.01 pH / ± 0.1mV | |
Accuracy | ± 0.1 pH / ± 5mV | |
ORP | Range | ± 2000mV |
Resolution | 0.1mV | |
Accuracy | ± 5mV | |
Temperature | Range | -5°C – +50°C (23°F – 122°F) |
Resolution | 0.01° C/F | |
Accuracy | ± 0.5° C |
Depth ** | Range | 0-60m (AP-2000 & AP-5000), 0 – 100m (AP-7000) |
Resolution | 0.01m | |
Accuracy | ±0.5% FS (AP-2000 & AP-5000), ±0.2%FS (AP-7000) | |
**Option on the AP-2000. Comes as standard on the AP-5000 & AP-7000 |
Ammonium | Range | 0 – 9,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L | |
Accuracy | ± 10% of reading | |
Ammonia† | Range | 0 – 9,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L | |
Accuracy | ± 10% of reading | |
Chloride | Range | 0 – 20,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 20,000 mg/L | |
Accuracy | ± 10% of reading | |
Flouride | Range | 0 – 1,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L | |
Accuracy | ± 10% of reading | |
Nitrate | Range | 0 – 30,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 30,000 mg/L | |
Accuracy | ± 10% of reading | |
Calcium | Range | 0 – 2,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 2,000 mg/L | |
Accuracy | ± 10% of reading | |
† Ammonium electrode required. Readings calculated from Ammonium, pH and temperature electrode values. |
Turbidity | Range | 0 – 3000 NTU |
Resolution | 2 Auto-range scales: 0.0 99.9 NTU, 100 – 3000 NTU | |
Repeatability | ± 2% of reading | |
Chlorophyll | Range | 0 – 500 µg/L (ppb) |
Resolution | 0.01 µg/L | |
Repeatability | ± 2% of reading | |
Phycocyanin (Freshwater Blue-Green Algae) | Range | 0 – 300,000 cells/mL |
Resolution | 1 cell/mL | |
Repeatability | ± 2% of reading | |
Phycoerythrin (Marine Blue-Green Algae) | Range | 0 – 200,000 cells/mL |
Resolution | 1 cell/mL | |
Repeatability | ± 2% of reading | |
Rhodamine WT Dye | Range | 0 – 500 µg/L (ppb) |
Resolution | 0.01 µg/L | |
Repeatability | ± 5% of reading | |
Fluorescein Dye | Range | 0 – 500 µg/L (ppb) |
Resolution | 0.01 µg/L | |
Repeatability | ± 5% of reading | |
Refined Oil | Range | 0 – 10,000 µg/L (ppb) |
Resolution | 0.1 µg/L | |
Repeatability | ± 2% of reading | |
CDOM/FDOM | Range | 0 – 2,000mg/L (ppm) |
Resolution | 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 2,000mg/L |
|
Repeatability | ± 10% of reading or 2ppm (whichever is greater) |
Hiện tại chưa có ý kiến đánh giá nào về sản phẩm. Hãy là người đầu tiên chia sẻ cảm nhận của bạn.