The Agilent PVG-500 / 500S(tungsten) / 502S(nickel) series of Pirani gauges provides the ultimate combination of state-of-art technology and ruggedness within a compact design. The PVG-500 series utilizes the most advanced digital Pirani technology combined with a stainless steel sensor design to meet the various needs of the market today.
Part Number | Description |
PVG500KF16 | PVG-500 Pirani gauge, KF17 |
PVG500KF16RS | Replacement sensor, PVG-500 Pirani gauge, tungsten filaments, KF17 |
PVG500KF16S | PVG-500 Pirani gauge, tungsten filaments, with setpoints, KF17 |
Technical Specifications
Measurement range |
5 x 10-4 to 1000 mbar (3.75 x 10-4 to 750 Torr) |
||||||
Accuracy (N2) |
Units |
||||||
% of reading |
+ 15% |
1 x 10-3 to 100 mbar (1 x 10-3 to 75 Torr) |
|||||
+ 50% |
5 x 10-4 to 1 x 10-3 mbar (3.75 x 10-4 to 1 x 10-3 Torr) |
||||||
+ 50% |
100 to 1000 mbar (75 to 750 Torr) |
||||||
Repeatability (Air) |
1 x 10-3 to 100 mbar (1 x 10-3 to 75 Torr) – % of reading +2% |
||||||
Output signal (measurement signal) |
Voltage range |
Measurement range |
|||||
0 to 10.3 V |
1.9 to 10.0 V |
||||||
Voltage vs. pressure |
1.286 V/decade, logarithmic |
||||||
Error signal |
0 to 0.5 V (filament rupture) |
||||||
Minimum loaded impedance |
10 kΩ, short-circuit proof |
||||||
Response time |
80 ms |
||||||
Adjustment |
One tactile switch for both ATM and HV adjustment |
||||||
Identification gauge |
27 kΩ, referenced to supply common |
||||||
Setpoint (2) |
|
PVG-500 |
PVG-500S, PVG-502S |
||||
Setting range |
None |
2 x 10-3 to 500 mbar (1.5 x 10-3 to 375 Torr) |
|||||
Hysteresis |
10% above lower threshold % of reading |
||||||
Relay contact |
30 VDC / 0.5 ADC floating |
||||||
Switching time |
|||||||
Supply voltage |
At gauge |
Ripple |
Power consumption |
||||
14 to 30 VDC |
≤1 Vpp |
≤1 W |
|||||
Electrical connection |
FCC 68/RJ45 appliance connector, 8 poles, male |
||||||
Sensor cable |
8 poles plus shielding |
||||||
Cable length |
≤ 100 meter (330 ft) |
||||||
Materials exposed to vacuum |
Glass, Ni, NiFe, tungsten (tungsten version) |
||||||
Admissible temperature |
Operation |
Storage |
Vacuum Connection |
||||
5 to 60 °C |
–20 to +65 °C |
80 °C |
Outline Drawing