Factors to Consider When Selecting a Stereo Microscope

15897 views - 07-02-2017, 9:47 am

Contents

 

  • Greenough and Cycloptic® principles
  • Criteria for selecting a stereo microscope
  • Magnification, zoom range and object field
  • Depth of field and numerical aperture
  • Ergonomics – people are very different
  • Illumination
  • Conclusion

 

 

Greenough and Cycloptic® principles

 

Binocular microscopes of the old days featured a simple lens system and the same design as traditional compound microscopes. These dissecting microscopes, as they were then known, were used primarily in biology for dissection purposes; there were no technical applications for them at the time. Around 1890, the American biologist and zoologist Horatio S. Greenough introduced a design principle which is still used today by all major manufacturers of optical instruments. Stereo microscopes based on the "Greenough principle" deliver genuine stereoscopic images of a very high quality.

 

In 1957 the American Optical Company introduced the modern stereo microscope design with a shared main objective and named it Cycloptic®. Its modern aluminum housing contained two parallel beam paths and the main objective, as well as a five-step magnification changer. This stereo microscope type, which was based on the telescope or CMO (Common Main Objective) principle, was adopted in addition to the Greenough type by all manufacturers and used for modular, high-performance instruments. Two years later, another American company, Bausch & Lomb, presented its StereoZoom® Greenough design with a groundbreaking innovation: a stepless magnification changer (zoom). Almost all of today’s designs are based on a zoom system.

 

Kính hiển vi soi nổi hiện đại Cycloptic® đầu tiên sử dụng nguyên lý của kính thiên văn

 

Fig. 1 (left): Cycloptic®, the first modern stereomicroscope based on the telescope principle.

Fig. 2a, b (right): Two basic stereo microscope principles. a: the telescope or CMO principle b: the Greenough principle.

 

 

Criteria for selecting a stereo microscope

 

Stereo microscopes are still based on the mentioned technical approaches – Greenough or CMO principle – today. Which other factors need to be considered? Four things need to be carefully assessed:

 

a)  What is the application?

b)  Which structures need to be observed, documented or visualized?

c)  How many people are using the microscope?

d)  What is the available budget for the solution?

Once the above factors are known it boils down to the following criteria

1. Magnification, zoom range and object field

2. Depth of field and numerical aperture

3. Optical quality and working distance

4. Ergonomics

5. Illumination

 

 

Magnification, zoom range and object field

 

The total magnification of stereo microscopes is the combined magnification of the magnification changer, the objective and the eyepieces.

 

The magnification changer or zoom body

 

Like a magnifying glass, the magnification changer consists of optical lenses that can be used to change the magnification of the instrument. Changing the position of the magnification changer alters the degree to which the image is magnified. The degree to which the image is magnified is called the magnification factor. Modern stereo microscopes are able to provide up to 16x magnification (zoom body only) with a 20.5:1 zoom range and feature motorization or encoding to allow reliable measurements.

 

Next, the image is magnified further by the eyepieces. To find out the magnification of the object he or she is observing in the eyepieces, the user has to multiply the magnification factors of the magnification changer and the eyepieces.

 

For the sake of completeness, however, here is the formula:

 

Công thức tính độ phóng đại tổng cộng

 

MTOT VIS is the total magnification that we want to calculate. VIS stands for "visual".

is the level of the magnification changer.

ME is the magnification of the eyepiece.

MO is the magnification of the main objective (1x in case no supplementary lens is used in a Greenough System)

 

Object field

 

When looking into the eyepieces from the proper distance and with the interpupillary distance set correctly, a circular area called the object field is visible. The diameter of the object field changes depending on the magnification. In other words, a mathematical relationship exists between the magnification and the diameter of the object field. Eyepieces with 10x magnification provide a field number of 23. That means at a 1x magnification of the zoom body and the main objective the object field is 23 mm in size. At 3x magnification the object field is reduced to one third, i.e. the object field has a diameter of only 7.66 mm.

 

Depth of field and numerical aperture

 

In microscopy, depth of field is often seen as an empirical parameter. In practice it is determined by the correlation between numerical aperture, resolution and magnification. For the best possible visual impression, the adjustment facilities of modern microscopes produce an optimum balance between depth of field and resolution – two parameters which in theory are inversely correlated.

 

Practical values for visual depth of field

 

The author of the first publication on the subject of visibly experienced depth of field was Max Berek, who published the results of his extensive experiments as early as 1927. Berek’s formula gives practical values for visual depth of field and is therefore still used today.

 

In its simplified form, it is as follows:

 

Giá trị thực nghiệm của độ sâu trường quan sát

 

TVIS:  visually experienced depth of field

n:  Refractive index of the medium in which the object is situated. If the object is moved, the refractive index of the medium that forms the changing working distance is entered in the equation.

λ:  Wavelength of the light used, for white light, λ = 0.55 μm

NA:  Numerical aperture on the side of the object

MTOT VIS: Total visual magnification of the microscope

 

If in the above equation the total visual magnification is replaced by the relationship of useful magnification (MTOT VIS = 500 to 1,000 x NA), it can be seen that, to a first approximation, the depth of field is inversely proportional to the square of the numerical aperture.

 

Đường cong biểu diễn sự phụ thuộc của độ sâu trường quan sát như một hàm của khẩu độ số NA với bươc sóng λ

Fig. 3: Depth of field as a function of the NA for λ = 0.55 mm and n = 1.

 

Particularly at low magnifications, the depth of field can be significantly increased by stopping down, i.e. reducing the numerical aperture. This is normally done with the aperture diaphragm or a diaphragm on a conjugated plane. However, the smaller the numerical aperture, the lower the lateral resolution.

 

It is therefore a matter of finding the optimum balance of resolution and depth of field depending on the structure of the object. In the case of stereo microscopes it is often necessary to make a certain compromise in favor of higher depth of field, as the z-dimension of three-dimensional structures frequently demands it.

 

Even more depth of field – FusionOpticsTM

 

A sophisticated optical approach that cancels the correlation between resolution and depth of field in stereo microscopes is FusionOptics™. Here, one of the light paths provides one eye of the observer with an image of high resolution and low depth of field. Via the second light path, the other eye sees an image of the same object with low resolution and high depth of field. The human brain combines the two separate images into one optimal overall image that features both high resolution and high depth of field.

 

Another example illustrating the phenomenal capabilities of the human brain is the Greenough stereo microscope. Here, the object planes of the left and right light paths are at a slight angle to each other. In the overall image, the entire hatched area appears to be sharply focused, although this is not the case in either the left or the right image.

 

Tiêu diện vật của kính hiển vi soi nổi sử dụng hệ quang học Greenough với độ sâu trường quan sát

 

Fig. 4: Object planes of the Greenough stereo microscope with depth of field range.

 

 

Kính hiển vi soi nổi hiện đại trang bị hệ thống Zoom 20.5:1 với thệ thống quang học APO và FusionOptics™

 

Fig. 5: Modern stereo microscope featuring a 20.5:1 zoom range with APO corrected optics and FusionOptics™.

 

Optical quality

 

The optical quality for stereo microscopes is usually listed as Achro or Achromat (achromatic), and as Apo (apochromatic) for the highest degree of correction for spherical and chromatic aberrations. Field curvature corrections are abbreviated Plan, while PlanApo designates a combination of chromatic aberration and field curvature correction.

 

Achro, Achromat Achromatic aberration correction
Plan Flat field optical correction
PlanApo Apochromatic and flat field correction

 

 

What is a chromatic aberration?

 

In optical instruments such as stereo microscopes, a chromatic aberration is a type of distortion in which there is a failure of a lens to focus all colors to the same convergence point. It occurs because lenses have a different refractive index for different wavelengths of light (the dispersion of the lens). The refractive index decreases with increasing wavelength. The aim of a good optical design is to reduce or eliminate this effect completely.

 

An achromatic lens or achromat is a lens that is designed to limit the effects of chromatic and spherical aberration. Achromatic lenses are corrected to bring two wavelengths (typically red and blue) into focus in the same plane. These types of lenses or microscopes are used for tasks where color reproduction is not imperative and mainly geometrical characteristics are assessed. Apochromatic lenses on the other hand are designed to correct three wavelengths (red, green, and blue) and bring them into focus in the same plane.



Working distance

 

This is the distance between the objective front lens and the top of the specimen when the specimen is in focus. In most instances, the working distance of an objective decreases as magnification increases. In stereo microscopy, working distance is one of the most important criteria, since it has a direct impact on the usability of the microscope as a tool.

 

 

Ergonomics – people are very different

 

There are tall and short people and this makes instrument requirements a personal matter. For example, the existing height of a microscope equipped for a certain task with accessories and with a particular working distance may be quite unsuitable for the specific user. If the viewing height is too low, the observer will be forced to bend forward while working, resulting in muscular tension in the neck region. Ideally, therefore, the viewing height and the viewing angle of the microscope should be adjustable to the build of the user. In addition, a variable viewing height is the best way to prevent an entirely sedentary posture. It permits the observer to adopt a personal sitting posture and to change it periodically in accordance with the natural urge to shift around from time to time. It is true that the height of the chair can be altered so that a relaxed, slightly bent posture is substituted for the previous rigidly upright one, but this is not the best approach. It is much simpler and more comfortable to use a variable binocular tube in order to compensate for the height difference.

 

Thanks to the modular product approach, stereo microscopes with a CMO design offer many ways of tailoring the instrument to the user’s size or working habits and are therefore the preferred solution.

 

Khoảng cách làm việc

Fig. 6: Ergo tube – Relaxed body and head, arms comfortably supported, adequate space for the legs, good use of the chair.

 

 

Illumination

 

In stereo microscopy, illumination is the key that will bring all of the work to light. The correct illumination will allow the required structures to be visualized or perhaps new information about a sample to be discovered, just by changing the type of light. It is important that the illumination is matched correctly to the right microscope and the right application.

 

 

Hệ chiếu sáng

 

Fig. 7: Modern stereo microscope illumination systems are based on long-lasting LEDs and provide unique ways to integrate the solution into the overall microscope system. Highly integrated ring light with applied polarizer to reduce glare on the specimen.

 

 

Types of illumination

 

Incident

Incident light is used with primarily non-transparent specimens. The method of delivering this light (ring light, spots, etc.) will depend on texture of the specimen and the application requirements. Incident light is needed for all kinds of non-transparent specimens. Depending on the texture of a specimen and the goal of the results, an eclectic selection of incident illumination solutions is available. 

Transmitted light

Transmitted light is desirable for various kinds of transparent specimens ranging from biological samples such as model organisms to polymers.

Standard transmitted brightfield illumination

Standard transmitted brightfield illumination is used for all kind of transparent specimens with high contrast and sufficient color information.

Oblique transmitted illumination

This illumination technique is used for specimens that are nearly transparent and colorless. Due to the oblique position of the illumination a greater contrast and visual clarity of the specimen can be achieved.

Darkfield illumination

Darkfield observation in stereomicroscopy requires a specialized stand containing a re­flection mirror and light-shielding plate to direct an inverted hollow cone of illumination towards the specimen at oblique angles. The principle elements of darkfield illumination are the same for both stereomicroscopes and more conventional compound microscopes, which are often equipped with complex multi-lens condenser systems or condensers ha­ving specialized internal mirrors containing reflecting surfaces oriented at specific geo­metries.

Contrast method for clear, transparent specimens

Rottermann Contrast™ is a partial illumination technique that shows changes of the refrac­tive index as differences in brightness. Phase structures then typically appear as spatial, relief-type images like hills in positive relief contrast and as indentations in inverted relief contrast. This technique offers many variable views for extracting the maximum possible amount of information.
 
 

Conclusion

 

Careful assessment of the application requirement for the stereo microscope is the key element for lasting satisfaction of the user. Since it is the workhorse of the laboratory or the production department, decision makers need to ensure that they are able to tailor the instrument 100% to user requirements. This requires a microscopy solution provider who is able to cope with this demanding requirement.

 

 