Order Information
Part Number | Description |
X3900-68003 | Heater wire, 10 AWG, precut 196 inches, for HS-32/HS-35 diffusion pump |
K0377169 | O-ring kit, for NHS-35 diffusion pump |
F1749001 | Heater clamping plate, for NHS-35 diffusion pump |
K1857335 | Halo baffle, 35 inch for NHS-35 diffusion pump |
F6097301 | Sightglass service kit, for NHS-35 diffusion pump |
L6370001 | Heat shield, reusable, NHS-35 diffusion pump |
K9050005 | Thermostat, upper switch, set at 104 °C, for HS-32 and NHS-35 diffusion pump |
K9050006 | Thermostat, lower switch, set at 316 °C, for NHS-35 diffusion pump |
L9699001 | Heater insulator, NHS-35 diffusion pump |
78536002 | Gasket for NHS-35 Diffusion Pump, ISO 800 Inlet Flange o-ring |
647235248 | Heater, NHS-35 diffusion pump, all voltages, without leads |
Technical Specifications
Pumping Speed, Operating Range Pumping Speed, AVS 4.1 (1963) | 28,000 l/s air, 35,000 l/s He and H2 50,000 l/s Air |
Maximum Throughput | 25 T-l/s (33 mbar l/s) in operating range, 35 T-l/s (45 mbar-l/s) @ 0.01 torr |
Operating Range | 9 x 10-4 to < 5 x 10-8 torr (1.2 x 10-3 < 6.5 x 10-8 mbar) |
Maximum Forepressure | No Load – 0.55 torr (0.71 mbar), Full Load – 0.40 torr (0.52 mbar) |
Recommended Backing Pump | ≥ 300 cfm (510 m3/hr) |
Backstreaming Rate, Standard Cold Cap | |
Warmup Time | 60 minutes |
Cooldown Time | 180 minutes |
Fluid Charge | 3 U.S. gallons (11.3 liters) |
Electrical Requirements | 3 ph, 50/60 Hz, 240/415/480 VAC |
Pump Power | 24,000 watts |
Cooling Water Requirements | 4 gpm (800 l/hr) at 60-80°F (15-26°C) |
Recommended Fluid | DC-704 |
Outline Drawing
Hiện tại chưa có ý kiến đánh giá nào về sản phẩm. Hãy là người đầu tiên chia sẻ cảm nhận của bạn.